Anaesthesia units are highly complex devices requiring specific infrastructure provision to be able to operate. They need a scavenging system, pipped medical air supply, and oxygen and nitrous oxide sources which can be piped or from cylinders. In addition, these machines require trained operators and technicians to carry out preventative maintenance so the units can operate safely.
The patient is anesthetized by inspiring a mixture of O?, the vapor of a volatile liquid halogenated hydrocarbon anaesthetic, and, if necessary, N?O and other gases. Because normal breathing is routinely depressed by anaesthetic agents and by muscle relaxants administered in conjunction with them, respiratory assistance is also provided via either an automatic ventilator or by manual compression of the reservoir bag.
Anaesthesia units dispense a mixture of gases and vapours and vary the proportions to control a patient's level of consciousness and/or analgesia during surgical procedures. Anaesthesia units primarily perform the following four functions:
- Blend gas mixtures, in addition to O?, that include air or nitrous oxide (N?O) along with an anaesthetic vapor.
- Facilitate spontaneous, controlled, or assisted ventilation while using these gas mixtures.
- Provide oxygen (O?) to the patient
- Reduce, if not eliminate, anaesthesia-related risks to the patient and clinical staff.
Suitable for adult, paediatric and newborn patients.
The unit is mounted on a trolley with minimum of four (4) anti-static swivel castors.
A side rail for mounting accessories is included.
The unit has an adjustable patient-circuit support arm.
The unit is provided with a minimum of three (3) gas inlets (O?, N?O and Air).
A minimum of two of the castors are provided with breaks.
The unit is equipped with an upper shelf.
The cart has been proved with handles for manoeuvring.
Gas inlet connections are compliant with DISS or NIST (needs to be specified when ordering).
The unit are equipped with gas supply gauges with scales allowing easy readout.
All pipeline connections have diameter-indexed safety systems (DISSs), or another means of preventing connection of dangerous gases.
The unit is equipped with a minimum of four (4) analogue rotameters, two (2) for O?, one (1) for air and one (1) for N?O with programmed parameters visualization.
The unit is provided with a minimum of two (2) gas cylinder yokes (O?, and N?O).
Holders for bottles are to include a secure locking mechanism.
The minimum gas supply input pressure range should be between 2.8 – 6 bar (41 - 87 psi).
The O? and N?O flowmeters have a minimum range of 0.0 – 10 L/min, and a resolution of at least 0.2 L/min.
The unit has the ability to carry out self-diagnosis and integrity testing, including a compliance and leakage test.
The unit is equipped with a minimum of two (2) vaporizer slots with a Selectatec mounting system.
The vaporizers are provided with an interlock system, preventing the use of more than one vaporizer simultaneously.
One (1) Halothane vaporizer is included with the unit.
Externally supplied N?O, O?, and air mixture ratios are fully controllable. There are separate controls for the flow of: O?, N?0, and anaesthetic agents.
Flows and the mixture ratios determined from flowmeter settings are accurate to within ± 10% of set values or better.
The unit basic is equipped with a passive scavenging system and optional allow for an active scavenging system.
The unit is equipped with a soda lime CO? absorber.
The unit will switch back and forth between battery and mains operation in case of power failures. Recharging resumes automatically when extern power becomes available.
The unit is designed and constructed in such a way that it is able to withstand frequent disinfection with hospital-grade products.
The unit is equipped with a non-return and a three-way valve, including the connection tube.
A built-in rechargeable battery allows for an autonomy of at least one (1) hour.
Fuses are fitted on both live and neutral supply lines.
Power requirements: 100 - 240 Volts - 50/60 Hz (not necessarily in a single unit).
VENTILATOR
The unit should be able to provide a minimum guaranteed tidal volume of 20 – 1,500 mL.
The unit should be able to support a minimum ventilation frequency range between 1 – 100 bpm.
The unit should be equipped with a ventilator suitable for adult, paediatric, and newborn patients and minimally supporting the following ventilation modes:
- Stand-by Mode.
- Volume Controlled Mode.
- Manual Mode.
The unit supports a peak inspiratory flow range of 1 – 70 L/min.
The unit supports PEEP with a range of 3 – 30 cmH?O.
The unit supports an I/E ratio 4:1 to 1:10.
The unit supports pressure triggers in the range of 0 – 20 cmH?O.
The unit supports an inspiration pressure of 5 – 70 cmH?O.
Pressure gauges have a range of 0 – 100 cmH?O and an accuracy of: ±2.5 % or better.
Request a custom quote, full technical datasheet, or place a bulk / tender enquiry. Our specialists will get back to you within 1 business day with pricing, lead time, and export documentation.
As one of India's leading Anaesthesia machine,closed circuit,w/acc manufacturers, suppliers & exporters, we deliver tender-grade quality with documented compliance to institutions, laboratories & training facilities across 65+ countries.
Every Anaesthesia machine,closed circuit,w/acc unit is manufactured to international quality standards and tested before dispatch. Full QC documentation & calibration certificates included.
Established export channels across Africa, Asia, the Middle East & beyond. Complete shipping documentation — certificates of origin, packing lists, test certificates, customs paperwork.
Need a specific configuration of Anaesthesia machine,closed circuit,w/acc? We offer customisation for tender specifications, OEM branding, voltage variants & bulk institutional orders.
Technical consultation before purchase, installation guidance, operator training documentation, and responsive after-sales support — backed by our specialist engineering team.
Trusted by institutions, governments & industries worldwide for reliable, tender-documentation-ready supply.
We supply Anaesthesia machine,closed circuit,w/acc across seven major buyer segments worldwide:
Practical laboratory training & coursework experiments
STEM education & physics / chemistry / biology demos
Process quality, testing & calibration requirements
Clinical lab, diagnostics & pathology applications
Ministry, defence research & PSU lab infrastructure
UNICEF, WHO, UNDP & bilateral aid procurement
Skill development & technical training labs
Direct & sub-contractor tender fulfilment
Manufacturing scale & global reach
Documentation-ready supply for tender-grade procurement — certificates of origin, test certificates, calibration data & compliance matrices.
Laboratory Instrument India is a recognised manufacturer, supplier and exporter of Anaesthesia machine,closed circuit,w/acc from India. We supply this product across our complete export network — from individual institutional purchases to large-scale tender supply contracts.
Nigeria, Kenya, Ghana, Ethiopia, Tanzania, Uganda, South Africa, Egypt, Morocco, Algeria, Senegal, Cameroon
Bangladesh, Sri Lanka, Nepal, Myanmar, Vietnam, Indonesia, Philippines, Malaysia
UAE, Saudi Arabia, Oman, Qatar, Kuwait, Iraq, Jordan
All shipments include complete export documentation, customs paperwork & freight handling.
Common questions from buyers, procurement officers & technical evaluators about ordering, customisation, lead time & documentation.
Speak directly with a product specialist for technical clarifications, custom configurations & bulk tender pricing.